6.2 INSTALLATION SITE REQUIREMENTS
The site for installation for TN50 should meet the following requirements:
 A maximum ambient temperature of 22°C
 No direct sunlight into the cooling chamber
 Mains power supply within 3 m
 No air circulation (for example by air conditioning)
 Instrument may only be used indoor
 Main switch / fuses must be reached easily
 Hand wheel must be reached easily
 Relative humidity lower than 60%
 A minimum distance between wall and rear of the instrument of 10-20 cm
(8 – 10 " inches) should be guaranteed
High room temperatures and high humidity influence the cooling
i
capacity of the instrument and will lead to ice forming inside the
instrument.
6.2 INSTALLATION
For opening the chamber push the window to the top.
Remove the transport security underneath the specimen holder and position the waste
tray.
After installation, wait at least 4 hours before switching on the instrument. Failure to
comply may result in severe damage to the instrument.