ATMOS S 201 Thorax |
Quick Start Guide
Keylock
The keylock is acti vated automati cally if no keys are pressed within the
preset unlocked period. The unlocked period can be adjusted in the
setti ngs menu (between 1 and 20 min).
To manually lock the device, please press the Increase and Decrease keys
simultaneously, unti l the keylock symbol is displayed on the screen.
To unlock the device, please press the Increase and Decrease keys
simultaneously, unti l the keylock symbol is no longer displayed on the screen.
Alarms
Please promptly address all alarms indicated on the screen. Use the on-screen
troubleshooti ng suggesti ons or consult the Instructi ons For Use Manual.
To silence/suppress the alarm and return to the standard operati on screen,
please press the Enter key. The on-screen alarm indicator will conti nue to
be displayed unti l the cause of the alarm is resolved (i.e. target vacuum
achieved, batt ery recharged).
Therapy Progress
The fl ow, ti me and vacuum history can be recalled by pressing the Therapy
Progress key. Use the Enter key to switch between fl ow, ti me and vac
metrics. With the Increase and Decrease keys, therapy history setti ngs can
be adjusted. To return to the standard operati on screen, please press the
Enter key.
PLEASE NOTE: The actual fl ow is accurate to within 100 ml.
Used Canister Removal
CAUTION: Please wear gloves. Please dispose of used canister according
to facility and governmental regulati ons. Only disconnect the device if the
target value of 17 mbar has been reached.
Before removing the used canister, stop the vacuum by quickly pressing
the Start Pump/Stop Pump key. The on-screen therapy paused indicator
will conti nue to be displayed unti l the therapy has been restarted. Press
the blue canister release butt on, located on the right side of the device,
and swivel the used canister forward unti l the canister can easily be
removed from its hinges. Next, disconnect the used canister from the hose
set. The drain tube should remain in place. Seal the secreti on channel and
canister pop off valve with the provided protecti ve caps.