2.4 Floor Mounting Instructions

For floor mounting the manifold control panel
and headers, a manifold control panel stand kit
and an appropriate number of manifold header
stand kits should have been ordered separately.
One manifold control panel stand kit contains
all items needed to mount the control panel.
Depending on the number of header segments,
one or more header stand kits are required. A
header stand is required for each header segment.
Each header stand kit contains two vertical stands
and all necessary items to support one 2-cylinder
or 3-cylinder header segment on each side of the
manifold. Verify contents of stand kits (Figures 6
and 7):
1. Remove
manifold
shipping carton and place face up on cardboard
packaging insert.
2. Remove the control module enclosure by
releasing the two latches on each side and set
the cover aside.
3. Remove the four M8 hex head bolts from the
mounting bracket as shown in Figure 1.
NOTE: Return control panel to shipping carton if
necessary to protect it and keep it clean.
4. Only bracket will be needed at this time.
Set
manifold
control
installationlater.
5. Attach post base to one end of each 80-inch
channel as shown in Figure 8, Detail A. Use
two each 3/8" bolts, flat washers and channel
nuts per base. Tighten all four bolts.
6. Stand each channel / base side by side. Attach
manifold mounting bracket to both bases as
shown in Figure 8, Detail C. Use four each 3/8"
bolts, flat washers and channel nuts perbase.
Top of bracket should be level and positioned
77 1/2" above floor. Tighten all fourbolts.
4107 9013 94.03
control
module
from
panel
aside
for
7. Position entire assembly in desired manifold
mounting location. Mark location of eight base
mounting holes. Move assembly aside and
drill holes. Minimum 3/8" diameter mounting
boltsare recommended (by others).
8. Reposition assembly over holes and install
mounting hardware (by others). Tighten all
mounting bolts.
9. Hang manifold control module on tab of
mounting bracket (Figure 3). Using M8 hex
bolt, attach manifold to wall bracket through
slots in black plate (4 places).
10. Additional cartons containing mastervalves
and header assemblies will be shipled loose.
Attach the master valves and appropriate
header extensions to the manifold control
panel.Position cylinder inlet connections of
each header extension so they are pointing
out (away from the wall) and tighten header
extension (Figure 4).
11. Set header stand in a vertical position roughly
in alignment with two vertical control panel
stands.
Position header bracket against
channel and on bottom side of header as
shown in Figure 5. Attach bracket to channel as
shown in Figure 8, Detail B. Use two each 3/8"
bolts, flat washers and channel nuts. Tighten
all bolts.
12. Align header stand with two control panel
vertical stands, centering the header bracket
between two pigtail connections. Mark
locations of all four base mounting holes.
Move header stand aside and drill holes.
Minimum 3/8" diameter mounting bolts are
recommended (by others).
13. Reposition header stand over holes and install
mounting hardware (by others). Tighten all
mounting bolts.
2-6