Remote Transmitter - Remote Az20 Probe Cable Connections - ABB Endura AZ series User Manual

Integral and remote transmitter, combustion oxygen monitor
Table of Contents

Advertisement

Endura AZ series integral and remote transmitter
Combustion oxygen monitor
4.5 Remote Transmitter – Remote AZ20 Probe Cable Connections
Terminal
Tag ID
Number
1
H
2
H
3
SCN
4
T / C –
5
T / C+
6
ACJC
7
ACJC
8
CELL –
9
CELL +
10
PS2
11
P COM
12
PS1
13
SV2
14
SV COM
15
SV1
Table 4.3 Remote Transmitter – Remote AZ20 Probe Cable Connections
Fig. 4.6 Remote Transmitter – Remote AZ20 Probe Cable Connections
IM/AZ20E–EN Rev. F
Connection Type
Heater
Heater
Screen
Thermocouple (–ve)
Thermocouple (+ve)
PT1000 Cold Junction Compensation
PT1000 Cold Junction Compensation
Oxygen Input (–ve)
Oxygen Input (+ve)
Pressure Switch – Gas 2
Pressure Switch – Common
Pressure Switch – Gas 1
Solenoid Valve – Gas 2
Solenoid Valve – Common
Solenoid Valve – Gas 1
AutoCal Connection (Terminals 10 to 15)
10
11
P2
P COM
To Probe –
Refer to IM/AZ20P–EN for
Remote Probe Connection Details
4 Electrical Installation
12
13
14
P1
SV2
SV COM
SV1
2 H
1 H
Standard
Connection
(Terminals 1 to 9)
Cable Color Code
Brown
Blue
Screens
White
Green
Grey
Violet
Black
Red
White / Yellow
White / Black
White / Orange
White / Blue
White / Red
White / Green
15
9 CELL +
8 CELL –
7 ACJC
6 ACJC
5 TC +
4 TC –
3 SCN
17