Introduction
Chapter 1: Introduction

Blender Module (Factory-Installed Option)

The Blender Module provides blended oxygen from 21% to 100% to the
patient outlet on the Resuscitation Module.

Resuscitation Module (Factory-Installed Option)

The Resuscitation Modules contain the pneumatic outlets necessary for infant
resuscitation (see figure 1-7 on page 1-11). Controls and displays for the
module are located above and to the rear of the bassinet.
There are three variations of Resuscitation Modules: RM2001™ Gas Delivery
Module without AutoBreath™ Infant Resuscitator (available only in the U.S.),
Resuscitation Module without AutoBreath™ Infant Resuscitator, and
Resuscitation Module with AutoBreath™ Infant Resuscitator (not available in
the U.S.). The Resuscitation Module without AutoBreath™ Infant
Resuscitator, and the Resuscitation Module with AutoBreath™ Infant
Resuscitator have adjustable airway pressure relief but the RM2001™ Gas
Delivery Module without AutoBreath™ Infant Resuscitator does not.
NOTE:
When pressure in the hose delivering gas to the patient falls below -4 cm H
(-0.4 kPa), the one-way valve installed at the Patient Outlet connection opens.
In the unlikely failure of the gas supply, the one-way valve allows patient
inspiration.
RM2001™ Gas Delivery Module without AutoBreath™ Infant

Resuscitator

• Primary Outlet (RM2001™ Gas Delivery Module Only)—Use the
Primary Outlet to provide continuous gas flow to a breathing circuit at a
rate of 0 to 15 LPM of oxygen (as selected by the operator). The Flow Rate
control is a calibrated dial-type flow adjustment or flow meter. A fixed
internal safety relief valve is also provided and is always operable. This
valve provides redundant maximum pressure relief at 160.1 cm H
cm H
option, the FiO
Outlet (see figure 1-7 on page 1-11).
Page 1 - 10
O (15.7 kPa ± 1 kPa). For units equipped with a Blender Module
2
of the Blender Module setting is delivered to the Primary
2
Resuscitaire® Radiant Warmer Products Service Manual (82 990 45)
O
2
O ± 10
2