IMPORTANT: Apply sample only to the sample tip of the test strip. Do not apply
blood or control solution to the top of the test strip as this may result in an
inaccurate reading.
Incorrect
Correct
Hold the blood drop to the sample tip of the test strip until the check window is
completely full and until the meter begins to count down. If the check window does
not fill, do not add more blood to the test strip. You may get an E-5 message or an
inaccurate test result. Discard the strip and retest. Even if the meter begins to
countdown but check window does not fill, discard the strip and begin the test
again with a fresh test strip.
Incorrect
Correct
Code Number
Each package of test strips is printed with a code number (CODE), lot number
(LOT), unopened expiration date ( ) and control range (CTRL 1 and CTRL 2).