When the Totalizer readout exceeds 9 digits (inclusive of the decimal places), the display shows
two lower vertical bars on the left side to indicate overflow. The leftmost display segments will
look like the following patterns:
Low order digits in overflow:
Middle order digits in overflow:
High order digits in overflow:
The count shown on the display remains accurate until a second overflow occurs (2,000,000,000
is
reached). In this case the vertical bar does not reflect how many times overflow has occurred.
Resetting the Totalizer
Press A2 until "tot" is displayed. (You must be in "Totalizer" mode first.)
Hold A2 for 5 seconds. Four rotating zeroes start to appear.
Continue holding A2 until all zeros are completed and "tot" is displayed again. Then
release A2.
Note: If A2 is released before the rotating zeroes are completed, the resetting of the Totalizer is
ignored.
Pressure
FSMAN.6 081214
Hold until all 4 zeroes
are completed
Page 17