SECTION 3

3.1 Setup Menu

3.1.1 Choosing the Probe Connector

INSTRUMENT OPERATION
ion
The sens
3 Portable pH meter has a setup menu which
allows the analyst to choose options for connector choice, display
lock, temperature units, display resolution, automatic buffer
recognition, time, and date.
To access the Setup menu, press the
will be displayed, indicating that additional options are available
within the menu. Press the up or down arrow key to scroll to the
desired option, then press
When using the Setup menu, the screen will display a number in
the upper right numerical field, indicating which option is being
changed. Table 3 describes these options.
Table 3 sens
Setup
Number
1
Use BNC or 5-pin connector
2
Display lock (On or Off)
3
Temperature units (°C or °F)
4
Measurement resolution(0.0, 0.00, or 0.000)
5
Buffer auto recognition (6.86 or 7.00)
6
Time of day (24 hour clock)
7
Date (mm/dd)
8
Year (4 digits)
This setup ensures the potential from the appropriate electrode is
detected by the meter. Connect only one electrode to the meter.
1. From the reading mode, press
2. The
icon and the number
Setup
Change the connector choice by pressing
the selection between
3. When the desired option is selected, press
the reading mode.
23
SETUP
.
ENTER
3
ION
Setup Options
Setup Description
.
SETUP
(flashing) will be displayed.
1
and
.
BNC
5 pin
key. The arrow icons
; this toggles
ENTER
to return to
EXIT