Exiting the Calibration Routine

Exit the calibration routine by pressing the
mode. The instrument uses the last completed user-entered calibration or the
factory calibration if no user-entered calibration has been completed.

Deleting Calibration Points

1. Select the range containing user-entered calibration points. See
Ranges on page
2. Press and hold the
.
READ/ENTER
Note:
Calibration points can also be deleted in Edit mode.
3. Press
ZERO/SCROLL
Press
READ/ENTER
4. The left digit will flash. Press
appear after the numeral 9.)
5. Press
READ/ENTER
User-Entered Calibration,
2—4.
key until "USER", then "CAL" appears. Press
MENU
to select the point to delete (e.g., S0 or S1 or S2).
.
ZERO/SCROLL
to delete. Repeat for all points to be deleted.
2—25
key to return to measurement
MENU
until "dEL" appears. ("dEL" will
continued
Switching

Advertisement

Table of Contents
loading