4.5 Setting a Probe Calibration Reminder (cont.)

4.5.1 Entering Probe Information

Use the arrow keys to scroll to S/N and press
1.
Use the numeric keys to enter the serial number from the Calibration
2.
Data Sheet and press
hex nut of the moisture probe.
Use the arrow keys to scroll to the suffix box and press
3.
open the drop-down list.
Use the arrow keys to scroll to the desired suffix and press
4.
Use the arrow keys to scroll to Calibrated On and press
5.
Use the numeric and arrow keys to enter the most recent calibration
6.
date. When you have entered the desired month, day and year, press
.
[ENTER]
Note:
Refer to the calibration data sheet for the most recent calibration
date.
Use the arrow keys to move to Cal Reminder and press
7.
prompt asks you to select a time period for the next probe calibration.
The PM880 will display a calibration reminder for the probe based on
the last calibration date. GE recommends recalibrating a probe a
minimum of once a year.
Use the arrow keys to move to the desired selection and press
8.
Press
(OK) and then
[F3]
9.
The PM880 returns to taking measurements.
The calibration reminder is part of the site file; therefore, if you are not
planning to make any more changes to the site file, you should save the new
changes. Refer to Saving a New Site File on page 96.
Model PM880 Portable Hygrometer User's Manual
. The serial number is also scribed on the
[ENTER]
.
[MENU]
Chapter 4. Using Special Features
.
[ENTER]
[ENTER]
[ENTER]
[ENTER]
to
.
[ENTER]
.
. This
.
[ENTER]
51