Humidity
•
5 to 95% without condensation
EMI/ESD
•
CE class B Compliant (EN61326-1)
•
FCC 47-1 Part 15 Compliant
Drop test
In accordance with the Telcordia GR-196-CORE recommendations, the
2000 Platform resists the following test:
•
6 impacts dropped from a height of 1m on a pinwood floor of 5 cm
thickness (1 impact on each of its 6 sides, with power off).
Shocks
The 2000 Platform resists the following test:
•
3 shocks per axis along each of the 3 axes, with power off.
•
Impacts of 15g, 1/2 sine, duration 11 ms, at 10 second intervals.
Bumps
The 2000 Platform resists the following test:
•
1,000 bumps per axis along each of the 3 axes, with power off.
•
Jolts of 15g, 1/2 sine, duration 6 ms, at 1 second intervals.
Vibration
The 2000 Platform resists the following vibration tests:
User Manual
Chapter 14 Technical specifications
79000000203 Rev. 005
Environment
159