Quick Card
®
T-BERD
/MTS-5800 Modular Test Set
E4100 OTDR Module Optical Light Source option
This procedure describes how to use an E4100-series OTDR module as an Optical Light Source (OLS). If
optioned, the OTDR module can operate as an OLS or Optical Power Meter, but not both at the same
time.
Equipment Requirements:
• T-BERD/MTS-5800 equipped with the
following:
o Fiber Optics Software Release V17.22
or greater
o E4100 Series OTDR Module
o E41OTDRLS Light Source option
• Fiber optic cleaning and inspection tools
• Launch Cable (patch cord) with
connectors matching the OTDR port and
Fiber Under Test
• Optical Coupler to connect Launch Cable
to Fiber Under Test
The following information is required to complete the test:
• Type of Fiber (Multimode or Single Mode)
• Type of Connectors (SC UPC, SC APC, LC UPC, etc.)
• Wavelength(s) to be tested (850nm, 1300nm, 1310nm, 1550nm, etc.)
Fiber Inspection Guidelines:
• Use the VIAVI P5000i or FiberChek Probe
microscope to inspect both sides of every
connection being used (OTDR Port,
Launch Cable, bulkhead connectors,
patch cords, etc.)
• Focus the fiber on the screen. If dirty,
clean the connector.
• If it appears clean, run inspection test.
• If it fails, clean the fiber and re-run
inspection test. Repeat until it passes.
Figure 1: Equipment Requirements
Figure 2: Inspect Before You Connect (IBYC)
VIAVI Solutions

Advertisement

loading

Summary of Contents for Viavi T-BERD 5800

  • Page 1 • Type of Connectors (SC UPC, SC APC, LC UPC, etc.) • Wavelength(s) to be tested (850nm, 1300nm, 1310nm, 1550nm, etc.) Fiber Inspection Guidelines: • Use the VIAVI P5000i or FiberChek Probe microscope to inspect both sides of every connection being used (OTDR Port, Launch Cable, bulkhead connectors, patch cords, etc.)
  • Page 2 VIAVI Solutions Connect to Fiber Under Test (FUT): All fibers and connectors should be inspected and clean prior to connection, as described on page 1. The Light Source may be connected to the FUT via an optical patch panel (OPP) or an optical coupler: 1.
  • Page 3 T-BERD/MTS-5800 until all testing is complete. © 2018 VIAVI Solutions Inc. Contact Us +1 844 GO VIAVI To reach the VIAVI office nearest you, Product specifications and descriptions in this (+1 844 468 4284) visit viavisolutions.com/contacts.

This manual is also suitable for:

Mts-5800