Faults - WIKA OLS-F1 Operating Instructions Manual

Optoelectronic level switches for sanitary applications
Hide thumbs Also See for OLS-F1:
Table of Contents

Advertisement

Available languages

Available languages

6. Faults

6. Faults
The following table contains the most frequent causes of
faults and the necessary countermeasures.
Faults
No function
Despite a change in
level, no alteration in the
display and no switching
of the output
Instrument responds
inversely
CAUTION!
Physical injuries and damage to property and the
environment
If faults cannot be eliminated by means of the listed
measures, the instrument must be taken out of operation
immediately.
Ensure that pressure or signal is no longer present
and protect against accidental commissioning.
Contact the manufacturer.
If a return is needed, please follow the instructions
given in chapter 8.2 "Return".
WIKA operating instructions model OLS-F1
Causes
Failure in power supply Measure voltage, check
Glass tip dirty
Glass tip too close to
an opposite surface
Wrong switching
function (normally
open, normally closed)
Measures
cable or plug connection
Clean the glass tip
(introduce service
intervals)
Increase the distance,
change the mounting
location
Replace instrument
EN
15