Example report format change (cont'd)
When all changes are complete, highlight Save and press Set.
To exit the Report Format Editor and return to the Set Up/
Preset Program Menu, highlight Exit and press Set.
To Add or Delete a measurement:
Enter the Report Format Editor in the same manner as "To
change the position".
Use the Trackball to highlight the desired measurement name
(to delete or change) or highlight a blank area (to add).
To delete a name, type in all spaces.
.
To change a name, type over the current name.
To add a name, type in the new name.
NOTE: When adding or changing a measurement name, it must
contain 2 important elements:
S
S
Highlight Save and press Set to record all changes.
.
Highlight Exit and press Set to return to the Set Up/Preset
Program menu.
NOTE: If 2 measurement names are placed in the same
position, an error message appears.
"More than 1 Measurement Field Name at the SAME Ypos".
LOGIQ
500 Basic Users Manual
2233658–100 Rev. 0
Measurement type. For example, GS, BPD, CRL, HC.
Author's name. The author's name must appear in
parentheses, for example (Hadlock), (Hellman).