8 |
3. Select
Device Check
4. Select
Start
5. The device check is running automatically. Wait for the test phase to finish.
The device will announce the test result.
If the device check fails, check the message and, if necessary, the
connectors. If the problem persists, stop using the device and contact
customer service.

1.2. SELECTIOn and aPPLICaTIOn OF ThE ELECTrOdE BELT

1.2.1. Choose the right belt size
The application requires an EIT patient interface. This consists of an elastic
electrode belt (16 electrodes) and an attachable patient cable. In order to cover
a wide patient spectrum, there are five belt sizes, from 70 cm to 150 cm chest
circumference. Electrode belt and patient cable are marked in size and color.
Both the electrode belt and the patient cable are reusable and can be disinfected
with wipes.
Size table
.
and confirm with the Rotary Knob .
Size in cm
XXL
124 – 150
XL
106 – 127
L
92 – 110
M
80 – 96
S
70 – 85
Size in inch
49 – 59
42 – 50
36 – 43
31 – 38
28 – 33