AquaProbe FEA100 / FEA200
Electromagnetic flowmeter – insertion-type flow sensors
Fig. 3.3 Maximum Permissible Velocity for Different Pipe Sizes
It is important to add the external length from the fixing point to the insertion length. Failure to do this can
give incorrect information from the graphs, resulting in vortex shedding affecting AquaProbe.
Examples:
A 600 mm pipe with the probe mounted on the centre line has an insertion length of 300 mm.
A typical valve is approximately 250 mm high and the distance to the support point inside the probe
is approximately 100 mm therefore, in this example, the total effective length is 650 mm.
Max velocity at 650 mm is 3.6 m/s.
8
6.0
5.0
4.0
3.0
2.0
1.0
0
0
200
Fig. 3.4 Maximum Permissible Velocity for Different Insertion Lengths
12
Insertion
Length
Pipe Size in Inches
16
24
32
40
400 600 800 1000 1200 1400 1600 1800 2000
Effective Probe Length in mm
Effective
Probe
Length
48
56
64
72
80
3 Mechanical Installation
20.0
17.0
13.0
10.0
7.0
3.0
0
Traversing
OI/FEA100/200–EN Rev. D