6.10.4 Remote interlock connector on the supply unit

figure11.
Position of the remote interlock connector
Page 36
The remote interlock connection is located
on the rear side of the supply unit (
12V DC operating voltage). The remote inter-
lock connector is plugged into this connec-
tion.
Remote interlocks that are connected to the
room, the door or other locally fixed safety
interlocks can be connected to the remote
interlock connection. The laser beam path is
interrupted if the contact is open.
The overall length of the cable between the
two connecting pins of the remote interlock
connector should not exceed 10 m.
User Manual TCS SP5