COOK Medical Fusion OASIS Quick Reference Manual

COOK Medical Fusion OASIS Quick Reference Manual

One action stent introduction system

Advertisement

Quick Links

Fusion
OASIS
®
O N E A C T I O N S T E N T I N T R O D U C T I O N S Y S T E M
Tips
1) fusion Oasis is compatible with
cotton-Leung®, cotton-Huibregste® and
ST-2 Soehendra Tannenbaum®.
2) The tapered tip end of the stent or
sideholes (if any) must be positioned in
the common bile duct while the other
end remains in the duodenum.
Step 2
Backload the device onto pre-positioned wire
guide, ensuring the wire guide exits the IDE
port.
M E D I C A L
www.cookmedical.com
®
Step 1
Place the desired stent and positioning sleeve onto the device, making sure the distal flap of the
stent is covered.
Step 3
Unlock the wire guide from the Wire Guide
Locking Device and advance the device into
the endoscope accessory channel. Once the
IDE port is inside of the accessory channel,
relock the wire guide.
Q u i c k R e f e R e n c e g u i d e
Step 4
Advance the device in short increments until
the stent is inside of the accessory channel,
then slide the positioning sleeve over the
pushing catheter, keeping it clear of the
accessory channel. Continue advancing
the device into the appropriate duct.

Advertisement

loading
Need help?

Need help?

Do you have a question about the Fusion OASIS and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for COOK Medical Fusion OASIS

  • Page 1 O N E A C T I O N S T E N T I N T R O D U C T I O N S Y S T E M Tips Step 1 1) fusion Oasis is compatible with cotton-Leung®, cotton-Huibregste® and ST-2 Soehendra Tannenbaum®.
  • Page 2 Step 5 Step 6 Step 7 Fluoroscopically monitor the radiopaque band Maintain position of the guiding catheter and Unlock the wire guide from the Wire Guide on the distal end of the guiding catheter. When advance the stent into the desired position Locking Device.