Caution:
After setting "Function" item as "DMM", directly pressing "AUTO TEST" key
can start a DMM test.
Pay attention to the measurement range descriptions in "Reference" menu of
the tester to ensure not to connect the out-ranged signals.
If one auto test is started and a signal with its voltage (>2V) is detected by the
tester,
Resistance" can not be measured and with no readings on the displays.
If one auto test is started and "Loop Resistance" is less than 40KΩ, "Loop
Capacitance" and "Isolation Resistance" can not be measured.
In "Loop Resistance", "Loop Capacitance" and "Isolation Resistance"
testing mode, the test can not be performed if there is a signal with its voltage
(>2V) is detected in the line.
The results in "Distance" menu are described in Table 20.
Item
Measurement
Unit
Metric (mm): 0.4, 0.5, 0.6, 0.9
Line
Diameter
Imperial (AWG): 26, 24, 22, 19
Loop
Resistance or
Capacitance
Distance
72-0046-03A
DI STA N
CE]
R u n n i n g . P l e a s e v i e w r e s u l t s
TES T
DI S T
Fig 27: "Distance" results menu
"Loop
Resistance",
Option
Metric, Imperial
——
——
Table 20: Distance estimation
[
M ETR IC ]
[0 . 4] m m
> 4 0. 0 0k Ω
> 1 0 . 0 0 k m
R EF ER E
"Loop
Capacitance"
Description
Unit for the measurement figures.
Refer to Table 21 and 22.
Same as the readings in "Test Result"
menu.
Distance=Loop Resistance (Capacitance) /
Resistance (Capacitance) constants
4 Navigating the Displays
and
"Isolation
49