Overview And Intrinsically Safe Data; Device Identification - Siemens SITRANS F M MAGFLO Operating Manual

Electromagnetic flowmeters transmitter type mag 6000 19" & safety barrier [eex ia/ib] iib sensor type mag 1100 ex & mag 3100 ex
Hide thumbs Also See for SITRANS F M MAGFLO:
Table of Contents

Advertisement

® ® ® ® ®
SITRANS F M MAGFLO
2.5 Overview and
intrinsically safe data
MAG 6000 19" IP 20
& IP 65 [EEx ia] IIB
MAG 1100 & MAG 3100
EEx ia IIB T3...T6

2.6 Device identification

Sensor MAG 1100 Ex or
MAG 3100 Ex label
8
2. Installation
Can only be installed in safe area!
All MAG 1100 Ex and MAG 3100 Ex sensors
have the following ratings and input parameters:
IS data sensor
MAG 1100 DN 6 - 100
MAG 3100 DN 15 - 300 Ex ib
Terminals
85-86
MAG sensor
coil
U
28 V
i
I
140 mA
i
P
2 W
i
L
2 mH
i
C
50 nF
i
SFIDK.PS.026.C2.02
Specifications:
Supply: 115-230 V or 24 V
Ambient temperature: −20 to 50°C
Enclosure: IP 20 or IP 65
IS data transmitter
Terminals
MAG barriers
U
o
I
o
P
o
L
o
C
o
Category 2 equipment
Sensors may be installed in zone 1 and zone 2.
Sensors intrinsically safe data
See table below.
MAG 3100 DN 350 - 2000 Ex e ia
82-83
Terminals
electrode
MAG sensor
10 V
U
i
50 mA
I
i
0.5 W
P
i
20 mH
L
i
50 nF
C
i
85-86
82-83
ib coil
ia electrode
28 V
9.3 V
138 mA
40 mA
0.4 W
4 mH
23 mH
100 nF
500 nF
85-86
82-83
coil
electrode
-
10 V
-
50 mA
-
0.5 W
-
20 mH
-
50 nF

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents