(2) Open the distribution section inner cover.
Distribution Section Inner Cover
(3) Align the "
" marks on the pipe slider and the
distribution unit.
"
" Marks
(4) Close the distribution section inner cover.
(5) Close the magazine door.
4
Select "No change" for "Cut interval".
NOTE:
When paper overlap feed has been occurred,
select "Change" for "Cut interval". If higher level is
selected (from level 1 to 3), the cut interval is
increased and the processing performance is
decreased.
5
Click the [Close] button.
• Operation returns to the menu screen.
9.2.21 Pump Output Measurement/
[Selection]
"4 Setup and Maintenance" - "6 Processor
Adjustment/Maintenance" - "1 Pump Output
Measurement/Setting"
[Procedure]
Perform only PSR replenisher pump output
measurement.
L2216
Pipe Slider
1
2
3
L21157
4
Setting (461)
Prepare the measuring cylinder (200ml).
Open the water supply port door.
Loosen the screw and remove the water supply
port inner cover.
Screw
Water Supply
Port Inner Cover
Click the [Start] button for "PSR".
FMPC
L21055
129