7. Troubleshooting
Error
CF15 Water
Reservoir Overflow
CF16 Ambient
Temperature Error
CF17 Air Drying
Error
CF21 Dosing Pump
Failure
CF22 Air
Temperature Error.
CF23 Top wash arm
low RPM
CF24 Bottom wash
arm low RPM
CF25 Reference
voltage Error
What it Means
Water reservoir overflow switch
triggered.
Unit is too hot or has overheated.
Drying system overheated.
Chemical dispensing system
error.
Faulty sensor.
Top wash arm speed too low.
Bottom wash arm speed too low.
System failure.
Page 39
What to Do
• Restart program.
• Call for service.
• The room or
enclosure is too
warm and not
allowing the unit to
adequately cool.
• If message persists,
call for service.
• Check HEPA filter
• Check if air exhaust
is blocked
• If message persists,
call for service.
• Tighten detergent
cap.
• Check for detergent
leak.
• If message persists,
call for service.
• Call for service.
• Check that load is
not blocking wash
arm.
• Check all holes in
wash arms for debris.
• If message persists,
call for service.
• Check that load is
not blocking wash
arm.
• Check all holes in
wash arms for debris.
• If message persists,
call for service.
• Call for service.