Applications
Figure 3 -4 .
5.
Read the percent ratio directly from the graduations on the
vertical centerline, referring to the 0% and 100% percentage
marks on the left edge of the graticule (1 minor division
equals 4% for a 5-division display).
TIME MEASUREMENTS
Tim e Duration
To measure time between two points on a waveform, use the
following procedure:
1. Apply the signal to either the CHI or the CH2 input connec
tor and set the VERTICAL MODE selector to display the chan
nel used.
2. Adjust the TRIGGER LEVEL control to obtain a stable display.
3. Set the SEC/DIV control to display between one and two
complete repetitions of the waveform.
DIV variable control is in the CAL detent.
3-8
READ PERCENT
RATIO (30%)
Determining voltage ratio.
6 7 1 7 -0 7
Check that the SEC/
2205 Operator's