The operating parameters menu contains the High Limit and Stir Speed parameters.
9.8.1.1

High Limit

The High Limit Parameter adjusts the upper set-point temperature. The factory default
and maximum temperature are set to 126°C. For safety, a user can ad just the HL down
so the maximum temperature set-point is restricted.
HL
Press "SET" to enable adjustment of HL
HL
H=126
Adjust the HL parameter using "UP" or "DOWN"
H=90
Press "SET" to accept the new temperature limit.
9.8.1.2

Stir Speed

The Stir Speed parameter adjusts stirrer motor speed. The factory default is 20.
Str SP
0
To change the stir speed press "UP" or "DOWN".
16
Press "SET" to accept the new Stir Speed.
The stir motor speed needs to be varied for best stability. Table 3 on page 25 shows
nominal settings for several fluids.
9.8.2

Serial Interface Parameters

The serial RS-232 interface parameters menu is indicated by,
SErIAL
High Limit parameter
Flashes "HL" and then displays the setting
Current HL setting
New HL setting
Flashes "Str Sp" and then displays the setting
Current Stir Speed setting
New Stir Speed setting
Serial RS-232 interface parameters menu
Controller Operation
Controller Configuration
39