Models FOB-NS:
Balance display
ò
(example)
ò
(example)
FOB-BA-e-1820
Operation
• Switch off the balance
• Hold down the ON/OFF key, simultaneously press
the UNIT key 3 times, release the ON/OFF key - an
internal numerical value is displayed.
Press the TARE key, "0" is displayed
Put adjustment weight centrically on the weighing
plate and press the TARE-key, "CAL" is displayed,
followed by the value of the put adjustment weight.
Take away adjustment weight
The balance changes into weighing mode.
Adjustment has now been completed successfully.
25