Models FOB-S
Balance display
ò
ò
ô
(example)
ô
(example)
24
Operation
Turn on balance by pressing the ON/OFF key
The balance carries out an autotest and changes
then into the weighing mode, the zero display „0.0"
appears
Press the UNIT-key for approx. 3 seconds, „CAL"
appears shortly, then „CAL 0" and after that „LoAd"
is displayed alternately with the value of the first
adjustment weight.
Place the first adjustment weight in the middle of the
weighing plate.
After a short time the value of the second
adjustment weight is displayed alternately with
„LoAd"
Remove the first adjustment weight from the
weighing pan and place the second adjustment
weight centrically on the weighing pan.
Adjustment will be carried out automatically after
stability control.
„donE" will be shortly displayed. Then the value of
the second adjustment weight will be displayed.
Adjustment has now been completed successfully.
The balance is now in weighing mode.