Model 60i Quick Start Guide Multi-Gas Analyzer Part Number 106474-00 23Jul2009 Example of a Direct Extractive Continuous Emissions Monitoring System (CEMS)
Pump Conditioner Numbered Steps Described in Table 2-1 Numbered Steps Described in Table 1 Figure 1. Rear Panel Plumbing Connections for 60i Analyzer and 61i Calibrator Table 1. Connecting Gas Lines Step Action Connect the sample line from the chiller/bypass pump to the SAMPLE bulkhead on the rear panel.
Page 3
10 feet. Verify that there is no restriction in this line. If calibration gases will not be introduced through the sample probe, Note see the “Calibration” chapter in the “Model 60i Instruction Manual” for alternative plumbing arrangement ▲ Step 1...
Page 4
Figure 3. Model 60i Plumbing Configuration Model 60i Quick Start Guide Thermo Fisher Scientific...
D-Sub connector on one side and a series of screw terminals on the other. Refer to the “Installation” chapter in the “Model 60i Instruction Manual” for detailed information on I/O configuration. Refer to Figure 5 and Table 2 for information on analog outputs and the most commonly used digital inputs.
Page 6
Analog Out #5 Conc. Analog Out #6 None 16,18,19,35,37 2,4,6,8,10,12 Signal Grounds Power Fail (NC) Power Fail (Com) Power Fail (NO) TTL Input 1 Turn Pump Off TTL Input 2 Trigger Blow Back Model 60i Quick Start Guide Thermo Fisher Scientific...
OFF and be sure to verify that the local power matches the label on the back of the analyzer. WARNING The Model 60i is supplied with a three-wire grounded power cord. Under no circumstances should this grounding system be defeated.
Set DHCP to OFF when changing the IP address. If DHCP is ON, ● the instrument will respond with “Not Settable if DHCP is ON.” For more information on DHCP, see the “Operation” chapter in the “Model 60i Instruction Manual.” Thermo Fisher Scientific Model 60i Quick Start Guide...
Page 10
Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific Air Quality Instruments 27 Forge Parkway Franklin, MA 02038 1-508-520-0430 www.thermo.com/aqi Model 60i Quick Start Guide Thermo Fisher Scientific...
Internal cables not connected Check that all internal cables are connected properly. Cycle AC power to Alarm - I/O properly instrument. Exp Status Defective board If still alarming, change board. Alarm - Motherboard Status Thermo Fisher Scientific Model 60i Quick Start Guide...
Page 12
Test Analog Outputs Set Analog Lock Percent Instrument Config Analog Input Cal Contact Info Display Pixel Test Raw Data Display Valve Leak Test 60i Pump Control Setup Dewpoint Span Cal Multipoint Lin Setup Model 60i Quick Start Guide Thermo Fisher Scientific...