Maintenance and Troubleshooting

1. Always keep the connector ports of the power meter clean.
2. Use the regulated optical connector for testing.
3. Use the adapters supplied with the kit only
4. Connect/disconnect fi ber connectors/adapters carefully to avoid
scratches on the port of the power meter.
5. Clean the optical port of power meter regularly. Clean using
cotton swabs supplied and isopropyl alcohol as directed.
6. When using AC adapter, ensure power supply is within the
required voltage range.
7. Remove the batteries when light source not in use for extended
periods of time.

Troubleshooting

Issue
Faint screen display
Inaccurate
Measurements
1. Ensure the connector is clean before testing.
2. Only use the supplied adapter.
3. Do not look into the laser when unit is on.
4. Charge the batteries before use. Do not charge in unit.
5. Cover optical port with dust-proof cap when not in operation.
6. Clean the optical port of the power meter regularly.
Handheld Optical Power Meter
Possible Reason
Low battery power
(A) Optical
Connector is not
clean
(B) Incorrect fi ber
connection
Warning
40
Solution
Charge or replace
the battery.
(A) Clean optical
connectors
(B) Reconnect the
fi ber