Calibrating Wavelength

1. Press
to select wavelength (dispalyed on top left of the
screen). 1310nm is the default value.
2. For testing, the wavelength value must be the same as the value
on the matched light source.
3. Multimode testing: 850nm or 1300nm
4. Single mode testing: 1310nm or 1550nm

Automatic Wavelength Detection

1. Press and hold
(pictured above). B/L set light will also illuminate. This indicates
that Wavelength Auto Identifi cation mode is active. Wavelength
will be detected based on the settings of the matched light
source.
2. To exit Wavelngth Auto ID mode, press and hold
disappears. B/L set light will stay illuminated until
3. For information on confi guring these settings with light source,
see next page.
Handheld Optical Power Meter

Wavelength

until --AU appears in the upper right corner
30
until --AU
is pressed.