Download Print this page

Test Port - Keysight Technologies PNA-L N5230A Configuration Manual

Pna series microwave network analyzers

Advertisement

– Add receiver attenuators (Option 016)
An attenuator is added between each test port and its corre-
sponding receiver. Two 35 dB step attenuators are added to the
E8362/3/4B. Two 50 dB step attenuators are added to the E8361A
(see PNA Series Microwave Data Sheet literature number 5988-
7988EN for a basic block diagram).
– Pulsed-RF measurement capability (Option H08)
Provides software to set up and control pulsed-RF measurements
with point-in-pulse capability. The software sets the coefficient of
the PNA's digital-IF filter to null out unwanted spectral components,
enables the IF gates provided with IF access (Option H11), and
controls selected Keysight pulse generators. It can be run on the
PNA or an external computer. A ".dll" file containing the IF-filter
algorithms is included for automated pulsed-RF testing. The pulsed
application is configured to work with the Keysight 81110A series
pulse generator.
For more detailed information regarding pulsed measurement
capabilities with the microwave PNA refer to the Keysight Web site
www.keysight.com/find/pna and download the PNA Series MW
Network Analyzers Configuration Guide for Pulsed Measurements,
literature number 5988-9833EN.
– IF access (Option H11)
Provides hardware to enable antenna, point-in pulse, and broad-
band millimeter-wave measurements to 110 GHz. For each of the
MW PNA's measurement receivers, IF gates (enabled with pulsed
measurement capability, Option H08) and external IF inputs are
added. In addition, access to the PNA's internal RF and LO source
is provided for remote mixing applications. For basic antenna
measurements, only Option H11 is necessary. Pulsed antenna
applications also require the pulsed measurement capability (Option
H08). Broadband measurements to 110 GHz, also requires an
N5260A millimeter-wave test set controller.
Note: Use external IF access for up to 20 dB more sensitivity when
making antenna measurements with a remote mixing configuration.
Add Option H08 (Pulsed-RF Measurement Capability) to enable
advanced pulsed measurements. Or upgrade to a broadband (10
MHz to 110 GHz) VNA system simply by purchasing an N5260A
controller test set with test heads (Option 110, 120, or 130).
1.
UNL* does not include bias-tees. Only includes source attenuators.
2.
Up to 67 GHz.
3.
The 5063-9237 kit assumes you have the standard handles shipped with the
instrument.
If you do not have handles, order a 5063-9224 kit.
– High-power test set (Option H85)
This configuration combines options that are often necessary for
high power measurements (UNL*
difference between ordering Option H85 versus a combination
of the options listed above is the source attenuator option UNL.
Standard UNL includes two source attenuators and two bias-tees.
Option H85 includes the two source attenuators, but not the bias-
2
tees, as the bias-tees are the power-limiting factor in the network
analyzer test set. The maximum power at the test port is +43 dBm
(<20 GHz), and +40 dBm (>20 GHz).
Option 080, frequency-offset mode, is included in Option H85
because it manages the phase-locking internally (instead of
depending on the R1 receiver). So if you need to use external com-
ponents in the path of the R1 receiver, it makes the measurements
simpler and more robust.
– Rack mount kit without handles (Option 1CM)
Adds a rack mount (5063-9217) and rail kit (E3663AC) for use
without handles.
– Rack mount kit with handles (Option 1CP)
Adds a rack mount (5063-9237)
standard supplied handles.
Documentation
PNA Online Help system is available within PNA instruments
in the following languages: English, German, Spanish,
French, Japanese, and Chinese.
The PNA Service Guide and Online Help are available on the
Web:
1
3
http://na.tm.keysight.com/pna
, 014, 016, 080, 081). The only
and rail kit (E3663AC) for use with
17

Advertisement

loading