V.A.C.® Therapy Battery Critical Alarm
Medium Priority Alarm - This alarm screen appears approximately 30 minutes before the battery power level is too low to support
continued operation of the V.A.C.ULTA™ Therapy Unit. This alarm will be accompanied by a repeating audible tone.
To resolve this alarm:
1.
Select
two minutes during troubleshooting.
2.
Connect the therapy unit to a wall outlet using KCI supplied
power supply to recharge battery. An amber light at the
bottom of the touch screen and a battery charge icon
indicates the unit is charging. Refer to the Charge Battery
section of this manual (page 23) for more information.
Once the V.A.C.ULTA™ Therapy Unit is plugged
into a wall outlet, the
screen will automatically clear.
3.
If the
not automatically clear, select
return to the
V.A.C.® Therapy continues, however, if this
alarm is not resolved within approximately
thirty minutes, therapy will be interrupted.
4.
Ensure therapy is ON by checking the
status bar (page 85). If not, select
Stop
The V.A.C.ULTA™ Therapy must be plugged into
a wall outlet in order to continue therapy.
Alarm logs and settings are not lost in the case
of a total power loss or if the unit is power
cycled (turned off then back on).
Certain KCI Dressings must be replaced with
an alternate dressing if therapy is interrupted
or off for more than two hours. See the
Safety Information Sheet provided with the
individual dressing for further information.
Audio Pause
to silence alarm for
Battery Critical Alarm
Battery Critical Alarm
screen does
Reset
Home
screen.
Start /
to restart therapy.