Battery Critical Alarm (Medium Priority)

This alarm appears about one hour before the battery power runs out. It is indicated by a
repeating audible tone and one flashing yellow LED.
Press and release
Do not press and hold.
Therapy will continue during this alarm. However, if the unit is not
plugged in within one hour, therapy will turn off.
To fix this alarm, recharge the battery.
1. Connect the therapy unit to a wall outlet using the power supply. The battery
charging light will glow amber as the battery charges. Refer to the
Instructions
2. Make sure therapy is on. The green LED on the power button will be lit.
3. If not, press and hold the
will turn the unit on and re-start therapy.
WARNING: If therapy stops or the unit shuts off for any reason, call
your doctor or nurse right away. Without active therapy, your dressing
will need to be replaced within two hours.
22
Audio Pause
chapter (page 12) for more information.
Power and Therapy On/Off
to silence the audible tone for two minutes.
button for three seconds. This
Battery Charging