V.A.C.® Therapy Canister Not Engaged Alarm
Medium Priority Alarm - This alarm screen appears when the canister is not fully inserted and / or properly latched. This alarm will
be accompanied by a repeating audible tone.
To resolve this alarm:
1.
Select
two minutes during troubleshooting.
2.
Remove the canister by pressing the
Canister Release
the unit.
3.
Inspect the canister and V.A.C.ULTA™ Therapy Unit to ensure
no foreign objects or debris interfere with the canister and
therapy unit's mating surfaces.
4.
Ensure both seals are present and seated completely (page
19). If seals are missing or damaged, contact KCI.
5.
Re-attach the canister to the V.A.C.ULTA™ Therapy Unit
ensuring that the canister is fully engaged and latched (page
32). An audible click indicates that the canister is properly
installed.
6.
Select
screen.
7.
Select
8.
If this alarm continues to appear, repeat steps 2 - 7 with a
new canister.
If alarm condition cannot be resolved, contact
KCI.
Audio Pause
to silence alarm for
button (page 18) on
Reset
Home
to return to the
Start / Stop
to restart therapy.