V.A.C.® Therapy Alarm Testing
1
2
3
4
5
1
2
3
4
5
0 0 : 0 0 : 4 5
0 0 : 0 0 : 2 4
0 0 : 0 0 : 1 0
0 0 : 0 0 : 0 8
0 0 : 0 0 : 0 5
1. Slide the untaped test canister into the side of the
V.A.C.ULTA™ Therapy Unit. An audible click indicates the
canister is fully seated .
2. Press the
Arrow
button displayed next to
3. After test is successfully completed remove the tubing cap
from the canister tubing.
4. Press the
Arrow
button displayed next to
5. After test is successfully completed install the tubing cap on
the canister tubing.
6. Press the
Arrow
button displayed next to
Blockage
Alarm.
7. After test is successfully completed remove the untaped
canister.
8. Press the
Arrow
button displayed next to
Engaged
Alarm.
9. After test is successfully completed slide the taped test
canister into the side of the V.A.C.ULTA™ Therapy Unit. An
audible click indicates the canister is fully seated.
10. Press the
Arrow
button displayed next to
Canister Full Alarm.
After tests are successfully completed a
show next to each item listed on the screen
(see figure at left).
11. Press the
Next
button to continue testing.
Pressure
Accuracy.
Leak
Alarm.
Canister Not
green check
should
25