Battery Critical Alarm

Medium Priority Alarm - This alarm screen appears about 30 minutes before the battery power runs out.
It is indicated by a repeating audible tone.
Battery Critical
Battery critically low; charge immediately
Audio Pause
Battery Low
Audio Pause
Reset
This alarm is fixed by recharging the battery.
1. Connect the therapy unit to a wall outlet using the
power supply. Refer to the
Instructions
Reset
On/Off
4. Press
Press
Audio Pause
to silence the audible tone.
The audible tone will come back on in two
minutes if the alarm condition has not been
fixed.
Therapy will continue during this alarm.
However, if the unit is not plugged in
within 30 minutes, therapy will turn off.
Battery Charging
section for more information.
2. Press
Reset
to return to the Patient Mode
Home
screen.
3. Make sure therapy is on. The
button will be green.
On / Off
to restart therapy if needed.
WARNING: If the therapy unit will be off
for more than two hours, call your doctor
or nurse right away. Without power to the
therapy unit, your dressing will need to be
replaced.
On / Off
21