System Error Alarm (At Power On) - Acelity V.A.C.ULTA 4 User Manual

Therapy system
Table of Contents

Advertisement

System Error Alarm (at Power On)

Medium Priority Alarm - This alarm screen appears when there is a system fault within the V.A.C.ULTA™ Therapy Unit while the unit
is powering on. "00000001" represents the diagnostic code of the system fault. This alarm will be accompanied by a repeating audible
tone.
To resolve this alarm:
1.
Record the Error Code number (00000001).
2.
Power the unit off and then on using the
Power button on the front of the unit
(page 18).
If alarm condition cannot be resolved, contact
KCI.
Certain KCI Dressings must be replaced with
an alternate dressing if therapy is interrupted
or off for more than two hours. See the
Safety Information Sheet provided with the
individual dressing for further information.
105

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents