Oxygen Purity Test - AirSep Onyx Instruction Manual

Psa oxygen generator
Table of Contents

Advertisement

6.4

Oxygen Purity Test

1)
Disconnect your application from the oxygen outlet.
2)
Set the unit's ON/OFF power switch to the ON position. (It takes
approximately five minutes for the oxygen purity to stabilize.)
3)
Verify that the product flow rate delivered by the unit matches the flow
capacity of the unit as specified in section 6.1 and does not exceed the
capacity of the unit.
4)
Connect a calibrated oxygen purity analyzer to the oxygen outlet.
5)
Take oxygen purity readings every 60 seconds until the analyzer shows the
same result for two consecutive readings.
6)
Disconnect the oxygen analyzer and reconnect your application.
6-2
Using the oxygen generator at flows higher than 15% above
those specified in Appendix A of this manual, will result in the
likely contamination of the molecular sieve beds. This damage
is not covered under the standard warranty.