AirSep
Corporation
5. Loosen and remove the solenoid base with a one-inch deep well
6. Install the rebuild kit, which contains all parts of the solenoid assembly
7.
8.3.3.2
If a solenoid valve coil does not operate, its corresponding green light on
the circuit board will not illuminate.
1. Set the ON/OFF switch to the OFF position, and unplug the power
2. Remove both sides and back panels.
3. Remove the red cap from the appropriate valve with a slotted head
4. Disconnect the solenoid leads, and lift off the solenoid coil.
5. Replace with the new coil.
6. Press the red cap back on top of the coil, and reconnect the solenoid
7. Reconnect the back and side panels.
8.3.4 Sieve Beds
Onyx/Onyx+/Onyx Ultra Instruction Manual
socket.
except the solenoid coil.
To reassemble the feed or waste valves, reverse the order of steps 1-
6, and test for leaks.
Solenoid Valve Coil Replacement
cord.
screwdriver.
leads.
Do not expose molecular sieve (contents of bed) to air for
an extended period of time. Prolonged exposure of
molecular sieve to the moisture in room air results in
contamination and permanent damage to the sieve
material. Keep all openings to the sieve beds sealed.
If replacement is necessary, you must replace both sieve
beds at the same time.