Psa oxygen generator
Table of Contents

Advertisement

®
AirSep
Corporation
compressor output results in less air for the sieve beds. Therefore, the
production of oxygen decreases.
Since this drop in oxygen production occurs over a very long period of
time, there is no regular preventative maintenance required on the
compressor. In the event of degradation in the performance of the unit,
the compressor should be replaced.
Sound Level
The condition of the compressor's bearings mainly determines its sound
level.
There are four bearings located within the compressor that allow the inner
components of the compressor to rotate. If the bearings wear to the point
that they become noisy, the compressor becomes noticeably loud and
needs servicing.
8.3.2.1
To remove the compressor assembly for exchange, follow the steps listed
below:
1. Set the unit's ON/OFF switch to the OFF position, and unplug the
2. Remove both sides and lower front panels.
3. Disconnect the suction tube.
4. Disconnect the blue compressor lead at the terminal strip and the
5. Disconnect the two leads to the capacitor and remove if necessary.
6. Disconnect the compression fitting for the heat exchanger located at
7. Remove the two screws that connect the compressor plate to the
8. Remove compressor from the plate by removing the four compressor
8.3.2.2
To install the compressor assembly, follow the steps listed below:
1. Perform the compressor removal procedure in reverse order.
2. Make sure to position the compressor's lead wires behind the braided
3. Leak test all connections.
Onyx/Onyx+/Onyx Ultra Instruction Manual
Compressor Assembly Removal
power cord.
brown lead at the temperature switch.
the bottom center of the compressor.
base of the unit, and slide out the compressor assembly.
bolts.
Compressor Assembly Installation
suction tube.
8-7