7.2
Cleaning the Enclosure
Use the following procedure to clean the enclosure:
1)
Shut down the oxygen generator as described in Section 6.3 — Shutdown.
2)
Disconnect the power cord from the electrical outlet.
3)
Use a dry, lint-free cloth on the enclosure, taking care to wipe the enclosure
clean.
7-2
To prevent electrical shock, shut down the oxygen generator
and disconnect the main power supply before you clean the
enclosure.
Dry cloth clean only. Do not apply liquid directly to the
enclosure or use any petroleum-based solvents or cleaning
agents.
AirSep
Onyx/Onyx+/Onyx Ultra Instruction Manual
®
Corporation