Maintenance; Cleaning The Gross Particle Filter - AirSep Onyx Instruction Manual

Psa oxygen generator
Table of Contents

Advertisement

®
AirSep
Corporation
7.0

Maintenance

The oxygen generator requires little maintenance. Use the following chart as a
guide to perform preventive maintenance at the required intervals.
Interval
Every week
Every month
* If product oxygen does not meet specifications, refer to Section 8.0 —
Troubleshooting.
7.1

Cleaning the Gross Particle Filter

Clean the gross particle filter on the back of the oxygen generator enclosure every
week, or more frequently if site conditions warrant. A clean filter allows the unit to
cool properly.
Use the following procedure to clean the gross particle filter:
1)
Shut down the oxygen generator as described in Section 6.3 — Shutdown.
2)
Remove the filter, and insert the second filter provided with the unit.
3)
Restart the unit as described in Section 6.1 — Start-up.
4)
Wash the filter removed from the unit in a solution of soap and warm water.
5)
Rinse the filter thoroughly and remove excess water with a soft, absorbent
towel.
6)
Allow the filter to dry, and then store it in a clean location. Alternate the clean
filter with the filter on the unit each time you perform this procedure.
Onyx/Onyx+/Onyx Ultra Instruction Manual
Clean gross particle filter. (Refer to Section 7.1 — Cleaning the
Gross Particle Filter.)
Make sure product oxygen remains within specifications in Appendix
A of this instruction manual.*
Clean enclosure. (Refer to Section 7.2 — Cleaning the Enclosure.)
Do not operate the unit without the gross particle filter in place.
Maintenance
7-1