Section 6: Leakage current and insulation resistance

Set up remote communications

You can run this application from the front panel or any of the supported communication interfaces for
the instrument (GPIB, USB, or ethernet).
The following figure shows the rear-panel connection locations for the remote communication
interfaces. For additional information about setting up remote communications, see
communication interfaces

Device connections

Depending on the device under test (DUT), the current measurement is typically very small, usually
<10 nA. Measuring leakage current and insulation resistance involves measuring very small values.
To get more accurate readings, connect the DUT to the Model 2450 rear panel with low-noise triaxial
cables.
Connect the DUT between the FORCE HI and FORCE LO terminals of the Model 2450.
The following figure shows schematic diagrams. One shows measuring the leakage current of a
capacitor. The other shows measuring the insulator resistance between two conductors of a coaxial
cable.
6-2
Model 2450 Interactive SourceMeter® Instrument User's Manual
(on page 3-1).
Figure 23: Model 2450 remote interface connections
Remote
2450-900-01 Rev. C / December 2013