Proceed with the following steps:
1. Press NEXT.
2. Field device has more status available
3. Press NEXT
4. Ignore next 50 occurrences of status?
5. Press YES
6. Change to MANual mode
7. Scroll to line 6 EXAMINE, press
8. Scroll down to 5 read status.
9. Read message.
10. Press OK.
11. Repeat OK to read all messages until the display returns to read status.
12. Scroll down to 6 clear status, press
13. If clear fault codes not completed appears, press OK and read the
message (Position Error, for example) or go to the troubleshooting guide.
14. Correct the problem (Is the air supply on?), and then go to clear status until
Clear Fault codes Completed appears.
15. Press OK.