Key Reference
Mode
Sets a value of the display offset on the current mea-
surement, and enable or disable the display offset.
The display offset is applied to the measurement sig-
nal after any math calculation.
• Average
Accesses the Average submenu.
Enables or disables the Average function.
Sets a value of the filter length and to set the average
count value to be Auto or Manual.
The power sensor automatically sets the number of
readings averaged together to satisfy most power
measurements averaging requirements. The number
of average readings depends on the resolution and
the power level settings the power sensor is currently
using.
Enables or disables step detection. In AUTO filter
mode, the average of the last four values entered into
the filter is compared to the entire filter average. If
the difference between the two averages is greater
than 12.5%, the digital filter is cleared. The filter will
then start storing new measurement data. This fea-
ture shortens the filter time required when the input
power changes substantially. Note that the result
appears to settle faster, although true settling to the
final value is unaffected.
• Meas Interval
Sets the interval time between two measurements.
The default value is 100 ms.
Key access: [MODE] > {Power Meter} > {Meas Setup}
82
Offset
Average
Avg Count
Step Detect