Using Functions
Using Markers and Pass/Fail Limit Lines
• Multiple limit segment
After setting a limit segment, the limit line is valid only
for your selected frequency/distance range and
amplitude range. A maximum of five limit segments
may be applied.
Setting up a limit line
Refer to the following example to add a limit line.
Example: adding limit 1
Step
1 Press [Limit] hardkey to access a
submenu of limit settings.
2 Press {Multi Limit} softkey to access
the multi limit segments submenu.
3 Press {Limit 1} softkey to access a
submenu of the limit 1.
4 Press {On} softkey to activate a limit
line.
4 Press [Edit] softkey to set the posi-
tion of a limit line.
5 Press 
{Prev segment/Next segment} 
to continue setting specifications for
a multi-limit trace.
42
Note
Use this feature if setting
multiple limit values on a
trace.
choosing the specific seg-
ment number.
This will set the limit line
at previously configured
limit value.
To adjust the limit line to
your specifications.
For multi limit traces.