Operation
4.12.3 Starting the Data Logging Process
After turnng ON data loggng and settng the
nterval and duraton, the analyzer wll pause
at the LOGGING SUMMARY screen, where the
currently selected nterval and duraton tme
perods are dsplayed.
Press ENT to start the combuston test and log-
gng process. At ths tme the Combuston Test
LOG screen wll appear, ndcatng that the ana-
lyzer s now performng a combuston test and the
data s beng stored n memory.
Note the followng:
• At the bottom of the screen, the memory
locaton where the current log entry s beng
saved s dsplayed. Ths number also repre-
sents the total number of test records that
are stored n memory.
• If there were prevously stored log entres,
each new loggng sesson wll start to store
data n the next avalable memory locaton.
• At no tme wll new data overwrte old data.
• The ENT button can be pressed whle loggng to vew the other Com-
buston Test screens.
4-14
LOGGING SUMMARY
Interval 1 min
Duration 2 hr
Press ENTER
to Start
MENU
F1
F2
LOG NATURAL GAS
O
4.0 %
2
CO
12 ppm
EFF
82.6 %
CO
9.5 %
2
T-STK
374 °F
T-AIR
68.0 °F
EA
21.3 %
CO(3)
13 ppm
Log Entry xx
F1
F2
Where: "xx" is the memory
location where the current
log entry is being saved.
Instruction 24-9448
F3
F3