Installation

Section 2
Use the proper lifting equipment when moving this unit. The unit's
shipping weight is 315 lbs. ((142.9kg).
To assure proper ventilation, allow a minimum of 4" of clearance between
the rear, top, and sides of the incubator and adjacent walls. If two or more
incubators are positioned side by side, allow a minimum of 8" between
cabinets. Adjust the front leveling feet of the incubator so that the front is
higher than the rear. This will assist in door closing. Choose a site free
from rapidly changing ambient temperature conditions.
Wall mount brackets are attached to the top rear of the cabinet. Remove
any packaging materials from the brackets. Rotate the brackets so they
point straight out from the back. Secure in place with the included
hardware. Attach brackets to wall.
Figure 2-1. Brackets During Shipping
Figure 2-2. Brackets During Use
The power cord is the mains disconnect. Make sure the outlet and power
cord are accessible at all times.
Caution The incubator should not be operated in an environment where
the ambient temperature exceeds 90°F as the compressor thermo-overload
will be tripped and will result in a wide band control cycle of
approximately ±4°C. Such cycling should not be misinterpreted as a
malfunction of the electronic controls.
s
Radiators, air-conditioning outlets, other ventilating system outlets, and
drafts can affect the operation of the incubator by a sudden inrush of air
that is at a temperature different than operating conditions.
Warning For personal safety, this apparatus must be properly grounded.
s
Thermo Scientific
Refrigerated Incubator
2-1