Section 6
Factory Options
Set Up the Incubator for
Communications (cont.)
6-4
Refrigerated Incubator
3. The display should have changed to the following. Again, the number
in the display might be different than 1. This selection is for Status
Display Type.
Listed are the numbered selections with explanations and examples of
the values displayed on your computer's monitor when the incubator
transmits information.
0 = Used for a raw, one-line status output suitable for importing into a
spreadsheet.
1 = Used for a multi-line format with English headings with continuous
screen.
2 = Used for the format as "1" but preceded by a clear screen command
used by terminals or computers running terminal software. This mode
overwrites the previous transmission.
With Status Display Type = 0:
10
-5.0
Where:
10 = Actual Temperature
-5.0 = Setpoint Temperature
0 = Normal
With Status Display Type = 1 or 2:
Actual Temperature: 10
Setpoint Temperature: -5.0
Alarm Status:
4. Allow the setup to time out.
The incubator is now ready to communicate with your computer.
0
Normal
LOW
HIGH