Transmitting data (USB interface)

52
9
Transmitting data (USB interface)
9.1

Options for data transmission

Via the USB interface you can transmit data to a PC. The following table shows
which data are transmitted to the interface in which way:
Data
Control
Current
Manual
measured val-
ues of all con-
nected
sensors
Automatic, at
intervals
Stored mea-
Manual
sured values
Calibration
Manual
records
Automatic
The following rule applies: With the exception of the menus, shortly
pressing the <F2>[USB output] key generally outputs the display
contents to the interface (displayed measured values, measure-
ment datasets, calibration records).
Operation / description
 With <F2>/[USB output].
 Simultaneously with every manual
storage process (see section 8.1
M
, page 46).
ANUAL STORAGE
 With <F2__>/[USB output].
Then you can set the transmission
interval.
 Simultaneously with every auto-
matic storage process (see section
8.2 A
UTOMATIC STORAGE AT INTER
, page 47).
VALS
 Displayed dataset with <F2>[USB
output] after calling up from the
storage.
 All datasets with the Output to USB
function.
(see section 8.3.1 E
MEASURED VALUE STORAGE
49).
 Calibration record with <F2>/[USB
output]
(see section 5.2.6 C
, page 32).
DATA
 At the end of a calibration proce-
dure.
ba75929e02
pH 7310
-
DITING THE
, page
ALIBRATION
10/2011