Miscellaneous commands

Miscellaneous commands are those commands that perform such functions as unlocking
calibration, saving calibration constants, locking out calibration, and programming date
parameters.
:CODE
(:CALibration:PROTected:CODE)
Purpose
Format
Parameter
Description
Note
Example
:COUNT?
(:CALibration:PROTected:COUNT?)
Purpose
Format
Response
Description
Example
To unlock calibration enabling you to perform the calibration proce-
dures.
:cal:prot:code '<password>'
Up to an 8-character string including letters and numbers.
The :CODE command sends the password/code and enables calibration
when performing these procedures via remote. The correct password
must be sent to the unit before sending any other calibration command.
The default remote password is KI002520.
The :CODE command should be sent only once before performing
calibration. Do not send :CODE before each calibration step.
To change the code, first send the present code, then send the new
code.
The password parameter must be enclosed in single quotes.
If you change the first two characters of the password to something
other than "KI", you will not be able to unlock calibration from the
front panel.
:CAL:PROT:CODE 'KI002520'
To request the number of times the Model 2520 has been calibrated.
:cal:prot:count?
Number of times calibrated.
The :COUNT? query may be used to determine the total number of
times the Model 2520 has been calibrated.
:CAL:PROT:COUNT?
Calibration Reference
Send default code of KI002520.
Request calibration count.
B-3